RoCoil TCA-5

The voltage across the output terminals is above 250mV RMS. Correct this by using a larger wire gauge or moving the TCA-5 closer to the meter.

No, clipping the output signal does not harm the TCA-5.

Each channel dissipates approximately 5 watts of heat at full power.

The TCA-5 has a peak internal current of 19 amps.
Yes. For example, over-ranging the TCA-5 on any single channel can burn it out. Ensure that the TCA-5 is not delivering more than 5 Amps RMS.

Not at this time.

The TCA-5 can be used on single-phase circuits, however there is currently no “1 channel model.”

The TCA-5 includes a 90 degree phase shifter. It can only be used with 333mV sensors to report current, NOT POWER.

Yes, the phase angle will be reported 0.6 degrees higher than actual, which is also about 0.6% low for a PF of 0.866.

FREE DOWNLOAD

Metering Project Guide

DENT BLOG

More of your questions answered.

Software Downloads

 
Misplaced the software disk or upgraded your computer? Download a current version of DENT software for your power meter or data logger here.

Technical Documents

 

Visit DENT’s comprehensive online library of DENT’s technical content including product manuals, specification sheets, special application information, and more.

Check Out What Our Customers Are Saying

Let Us Help You Find The Right Product For Your Needs

Do you need help choosing the right equipment for your project? Or maybe you need technical help on a DENT product you already own? Share details below and we will be in touch within one business day. Remember that product manuals and specifications can also be found in our Technical Library.